I.7 Cleaning
Be sure to remove spindle from instrument prior to cleaning. Severe instrument
damage may result if cleaned in place.
Instrument and Keypad: Clean with dry, non-abrasive cloth. Do not use solvents
or cleaners.
Immersed Components (spindles): Spindles are made of stainless steel. Clean with non-abra-
sive cloth and solvent appropriate for sample material that
is not aggressive to immersed components.
When cleaning, do not apply excessive force which may result in bending spindles.