Brookfield Engineering Labs., Inc. Page 24 Manual No. M/94-310F
Figure B14
CONE (00) CALIBRATION
VISCOSITY FSR TEMP S.STRESS S.RATE SPEED TIMER CONE SAMPLE
(POISE) (%) (Deg C) (D/CM
2
) (1/sec) (RPM) (SEC) No. No.
Operator:
Date:
Model/Serial#:
Fluid:
Figure B11
TEMP. = 25.0°C
Verify the calibrating temperature is correct and stable. If the value is incorrect, press STOP,
return to Step 1 and begin again.
Press the RUN key to calibrate the cone. The screen will display:
Figure B12
CALIBRATING (00)
When calibration is completed, rotation will stop and the screen will display:
Figure B13
CAL COMPLETE
Pressing the PRINT key sends the calibrating parameters to the printer port (Figure B14) and
returns the CAP to the main screen.